HB-9210

Sharper analysis, simpler reading, greater autonomy: the Hb9210
Premier makes the very most of the IFCC and NGSP certified method for HbA1 c analysis - with A.
Menarini Diagnostics' experience and know-how to power and back it up.
No Hb variant interference

The Boronate affinity method prevents any interference due to the main abnormal Hb, e.g. Hb Variants, Carbamylated Hb or Labile Fraction.
66 s/test output

The Hb921 O Premier is by far the fastest HPLC analyser not affected by Variant haemoglobins on the market today.
210-position autosampler

The Hb921 O Premier's great sampling capacity (21 O samples with no downtime) endows it with extended autonomy.
lntegrated calibration & controls

State-of-the-art integration of calibration and control plus the Hb9210. Premier's capacity to integrate extra hardware guarantee 100% reliable walkaway automation, even for heavy overnight routines.
User-friendly software and touchscreen

Test results come in a two-peaks chromatogram on an exclusive touchscreen, which allows a quicker, more intuitive interaction with the system, while the Hb9210. Premier's software ensures total traceability of the samples, QC, reagents, users and activities.
CV< 2%

High precision diagnosis.
Double reporting

All results appear both in DCCT and IFCC units.

Technical Specification  
Dimension 74 cm H, 54 cm W, 66 cm D
Weight 62 kg
Element of Analysis Quantification of glycated haemoglobin
Prinicple Boronate affinity high performance liquid chromatography
Minimum Readable Division Area percent = 0,01%
Minimum sample requirements 10 microliter whole blood, 5 microliter packed red blood cells
Dilution Whole blood 1 : 1 00, packed red blood cells 1 : 200
Sample Capacity 210 samples + stat position
Thoughput 66 seconds per injection
Analytical Column Boronate bonded to a porous polymer support gel
Column Temperature 54°C
Dectection LED wavelength detector, 413 + 2nm
Software lnstalled on hard drive of Hb9210 Premier CPU
Calibrators On board, Glycated Haemoglobin Calibrators (2 levels)
Controls On board, Glycated Haemoglobin Controls (2 levels)

 